The BD LSRFortessa™ and BD™ LSRII cell analysers are high end analytical flow cytometers and offer unmatched performance, capabilities, and flexibility, utilising a revolutionary optical design. The electronics digitise signals 10 million times per second in 16,384 discrete levels. This expands channel resolution and increases detection sensitivity. Electronic dead time is eliminated; more precise fluorescent measurements can be made improving linearity, compensation, and better quantification of fluorescence measurements.
The core lab LSR II and Fortessa are equipped with five fixed alignment lasers (488 nm, 405 nm, 633 nm, 561 nm and 355 nm) and BD High Throughput Sampler (HTS) option. The BD LSRFortessa™ and BD™ LSRII cell analysers use FACSDIVA software for acquisition and analysis.
Configuration of the BRC LSR II and Fortessa analysers
BRC SORP Fortessa and LSR II: 5 laser, 20 parameter SORP instrument (355 nm (nUV), 405 nm (violet), 488nm (blue), 561nm (Yellow-green), 633 nm (red)), allow simultaneous use of all lasers. Equipped with HTS option. Laser, detector and filter configuration 3 Blue/3 Red/5 Violet /2 nUV/ 5 Yellow-Green:
|
Laser
|
LP Mirror
|
BP Filter
|
Intended Dye
|
|
488 nm
Optional
|
550 nm
|
575/26 nm
|
PE
|
|
505 nm
|
530/30 nm
|
FITC
|
|
Blank
|
488/10 and pinhole
|
SSC
|
|
505 nm
|
525/50
|
FITC
|
|
495 nm
|
515/20 nm
|
EGFP
|
|
525 nm
|
545/35 nm
|
YFP
|
|
635 nm
|
670 LP
|
PerCP, PerCP-Cy5.5
|
|
561 nm
Optional
|
735 nm
|
780/60 nm
|
PE-Cy7
|
|
685 nm
|
710/50 nm
|
PE-Cy5.5
|
|
635 nm
|
660/20 nm
|
PE-Cy5
|
|
595 nm
|
610/20 nm
|
PE-TexasRed,mCherry
|
|
570 nm
|
585/15 nm
|
PE
|
|
750 nm
|
780/60 nm
|
PE-Cy7
|
|
635 nm
|
670/30 nm
|
PE-Cy5
|
|
633 nm
Optional
|
735 nm
|
780/60 nm
|
APC-Cy7
|
|
685 nm
|
710/50 nm
|
Alexa700
|
|
Blank
|
670/14
|
APC
|
|
750 nm
|
780/60 nm
|
APC-Cy7
|
|
690 nm
|
730/45 nm
|
Alexa700
|
|
Blank
|
670/30 nm
|
APC
|
|
405 nm
|
755 nm
|
780/60 nm
|
QDot 800
|
|
|
690 nm
|
710/50 nm
|
QDot 700
|
|
600 nm
|
610/20 nm
|
QDot 605
|
|
570 nm
|
585/42 nm
|
QDot 585
|
|
505 nm
|
525/50 nm
|
AmCyan
|
|
Blank
|
450/50 nm
|
Pacific Blue
|
|
351 nm
|
505 nm
|
530/30
|
Indo1
|
|
|
Blank
|
450/50 nm
|
DAPI/Hoechst
|
Note
Optional: The core laboratory owns these filters as well, but they are not routinely installed in the cytometer. If a user requires filters for an experiment, please inform FCL staff in advance.
Booking
The analysers are available 24 hours a day, 7 days a week for use by investigators except during weekly maintenance and repair periods. For details of these, see the guidelines page.
Before an investigator can book and use the equipment, they must attend a basic training session in order to obtain log-in credentials to the instruments and online booking calendar.
Book time on BD Fortessa
Book time on BD LSR II
When booking, please note the following:
-
The equipment is available in 15 minute time slots.
-
No more than 4 hours can be continuously booked during prime time usage (Monday - Friday, 12pm - 8pm).
-
No single user can book more than 8 hours of prime time usage per week.
-
Members of the BRC's Faculty of Translational Medicine will be given priority access.