The BD FACSAria™ II Cell Sorter is a benchtop high-speed flow cytometry cell sorter with a fixed-alignment cuvette flow cell and up to 5 air-cooled lasers. It allows for high-speed multicolour sorting with digital acquisition rates of up to 20,000 events per second.
The flow laboratory is equipped with two Aria sorters. Both can sort into two- and four-way bulk sorting devices for a variety of tube sizes or can sort using Automated Cell Deposition Unit (ACDU) to multi-well plates or microscope slides. The BD FACSAria™ II Cell Sorters use FACSDIVA software for acquisition and analysis.
Configurations of the BD FACSAria™ II Cell Sorters
BRC ARIA 1 (operated by advanced and trained users only):
4 laser, 10 parameter instrument (355 nm (nUV), 405 nm (violet), 488nm (blue), 633 nm (red)), either the UV or the violet laser can be used, not both simultaneously. ACDU installed. Laser, detector and filter configuration 6 Blue/2 Red/2 Violet or nUV:
|
Laser
|
LP Mirror
|
BP Filter
|
Intended Dye
|
|
488 nm
optional
|
735 nm
|
780/60 nm
|
PE-Cy7
|
|
655 nm
|
695/40 nm
|
PE-Cy5.5, PerCP Cy5.5
|
|
610 nm
|
616/23 nm
|
PE-TexasRed
|
|
556 nm
|
585/42 nm
|
PE
|
|
502 nm
|
530/30 nm
|
FITC
|
|
Blank
|
488/10
|
SSC
|
|
655 nm
|
675/20 nm
|
Pe-Cy5, PerCP
|
|
633 nm
|
735 nm
|
780/60 nm
|
APC-Cy7
|
|
Blank
|
660/20 nm
|
APC
|
|
405 nm
|
505 nm
|
530/30
|
AmCyan
|
|
Blank
|
450/50 nm
|
Pacific Blue
|
|
351 nm
|
505 nm
|
530/30
|
Indo1
|
|
Blank
|
450/50 nm
|
DAPI/Hoechst
|
Note
Optional: The core laboratory owns these filters as well, but they are not routinely installed in the cytometer. If a user requires them for an experiment, please inform flow cytometry laboratory staff in advance.
Using the BRC Aria 1
The machine is user operated and available 24/7.
Only advanced flow cytometry users can access to the self service BCFACSAria II™ Cell Sorter. To become an advanced user, applicants must be approved by the senior flow cytometry manager and complete of a BD Aria operator training course.
Advanced flow cytometry users are issued with a username and password, which allows access to the sorter and the online booking calendar.
Members of the BRC's Faculty of Translational Medicine will be given priority access followed by non-members on a first-come, first-serve basis.
Flow cytometry laboratory staff reserve the right to review and/or revoke access to the self-service ARIA if a user repeatedly fails to safely operate the sorter. FCL staff also reserve the right to review and/or alter the time booked on BRC ARIA1 to allow maximum access to the sorter for all users.
Users are responsible for filling and emptying of all fluidics and for instrument setup and cleaning.
Users prepare and bring their own sheath buffer.
Book time on BRC Aria 1
BRC ARIA 2 (Flow cytometry laboratory operated):
5 laser, 20 parameter SORP instrument (355 nm (nUV), 405 nm (violet), 488nm (blue), 561nm (Yellow-green), 633 nm (red)), allows simultaneous use of all lasers. ACDU and Aerosol Management system installed. Laser/Detector and filter configuration (3 Blue/3 Red/5 Violet /2 nUV/ 5 Yellow-Green):
|
Laser
|
LP Mirror
|
BP Filter
|
Intended Dye
|
|
488 nm
Optional
|
550 nm
|
575/26 nm
|
PE
|
|
505 nm
|
530/30 nm
|
FITC
|
|
Blank
|
4881/10 and pinhole
|
SSC
|
|
505 nm
|
525/50
|
FITC
|
|
495 nm
|
515/20 nm
|
EGFP
|
|
525 nm
|
545/35 nm
|
YFP
|
|
635 nm
|
670 LP
|
PerCP, PerCP-Cy5.5
|
|
561 nm
Optional
|
735 nm
|
780/60 nm
|
PE-Cy7
|
|
685 nm
|
710/50 nm
|
PE-Cy5.5
|
|
635 nm
|
660/20 nm
|
PE-Cy5
|
|
595 nm
|
610/20 nm
|
PE-TexasRed,mCherry
|
|
570 nm
|
585/15 nm
|
PE
|
|
750 nm
|
780/60 nm
|
PE-Cy7
|
|
635 nm
|
670/30 nm
|
PE-Cy5
|
|
633 nm
Optional
|
735 nm
|
780/60 nm
|
APC-Cy7
|
|
685 nm
|
710/50 nm
|
Alexa700
|
|
Blank
|
670/14
|
APC
|
|
750 nm
|
780/60 nm
|
APC-Cy7
|
|
690 nm
|
730/45 nm
|
Alexa700
|
|
Blank
|
670/30 nm
|
APC
|
|
405 nm
|
755 nm
|
780/60 nm
|
QDot 800
|
|
|
690 nm
|
710/50 nm
|
QDot 700
|
|
600 nm
|
610/20 nm
|
QDot 605
|
|
570 nm
|
585/42 nm
|
QDot 585
|
|
505 nm
|
525/50 nm
|
AmCyan
|
|
Blank
|
450/50 nm
|
Pacific Blue
|
|
351 nm
|
505 nm
|
530/30
|
Indo1
|
|
|
Blank
|
450/50 nm
|
DAPI/Hoechst
|
Note
Optional: The core laboratory owns these filters as well, but they are not routinely installed in the cytometer. If a user requires them for an experiment, please inform FCL staff in advance.
Booking the BRC Aria 2
The Aria 2 is exclusively operated by flow cytometry laboratory staff.
The instrument and operators are available for sorts Monday - Friday, 10am-5pm.
All sorts are run with sterile 1x PBS as sheath buffer. Users can provide their own sheath buffer if necessary but must inform the Flow Cytometry Laboratory at least 2 days in advance.
Members of the BRC's Faculty of Translational Medicine will be given priority access followed by non-members on a first-come, first-serve basis.
The flow cytometry laboratory staff will assign sorting times and strive to avoid large blocks of unused time to maximise access. Users will be able to view the booking calendar.
To book sorting time on this instrument, send completed booking form (word) by email to BRCFlowCore@gstt.nhs.uk.